64

65

General

In case of malfunction

Problem

Potential cause

Solution

The hearing
aid is dead

The hearing aid is
not turned on

Make sure the battery
drawer is closed

The earwire is
broken

Contact your hearing
care professional

The battery is dead
or does not work

Insert a new battery in
the hearing aid

The sound outlet
of the ear-tip is
blocked

Closed ear-tip: Clean
the blocked sound
bore (see page 30)
Open ear-tip: Replace
or clean the ear-tip
Custom ear-tip: Clean
the ear-tip (see page
58)

The NanoCare wax
guard is blocked

Change the wax guard

Problem

Potential cause

Solution

The hearing
aid volume is
not powerful
enough

The battery is
nearly exhausted

Insert a new battery in
the hearing aid

The sound outlet
of the ear-tip is
blocked

Closed ear-tip: Clean
the blocked sound
bore (see page 30)
Open ear-tip: Replace
or clean the ear-tip
Custom ear-tip: Clean
the ear-tip (see page
58)

The NanoCare wax
guard is blocked

Change the wax guard

Your ear is blocked
by earwax

Contact your ENT
doctor/physician

Your hearing may
have changed

Contact your hearing
care professional