Widex AK-X User Manual
Page 29
29
Changing the CeruSTOP wax guard
Wipe away any visible wax from the shell of your
hearing aid. The used wax guard is in the stainless
steel bushing around the sound outlet. With the
holder perpendicular to the face of the bushing, in-
sert the removal hook into the used wax guard so
that the shaft of the holder is touching the rim of
the wax guard, then slowly pull the wax guard
straight out. There is no need to dig or twist the
wax guard from the bushing.
1. Vent
opening
2. Sound outlet
3. Used wax
guard