OPERATORS MANUAL
Machine Type:
PD22I / PD22P
PD22 I P OM 1.01 EN
Ver 1.01
Date 24-08-2006
Page 10 of 21
6
DISPENSING WITH PD22
For optimal dispensing with the PD22, the following should be observed:
6.1
Vessel
placement
6.2
Tubes
6.3
Nature of fill media
6.4
Priming
tubes
6.5
Drip
6.6
Hard
feed
6.1 Vessel
Placement
In order to build up adequate pressure and reduce friction, it is recommendable to place the
vessel containing fill media at the same level as pump head or preferably above the pump head
level. Placing the vessel higher than pump head level provides positive product support and
may reduce the calibration interval. It is also recommended to place the vessel as close as
possible to the pump on suction side.
Flexicon
Pump Head
Preferred Placement
of container
Normal Placement
of container
Suction side
Pressure side
Fig. 6.1