Manual operation

Insert the mains transformer jack plug into the round socket on the pump back panel.

Pumps delivered with multi-plug mains adaptor: see pictures above for how to change plugs.

Switch power on at the mains socket. Power at the pump will be signalled by the illumination of a green
LED on the pump front panel.

The manual speed control potentiometer has three positions.

• The speed setting dial is calibrated in proportion of maximum speed from 0-10 which represents a 10:1

speed reduction ratio from the maximum speed of the pump. Turn the control knob clockwise to increase
the speed and counter-clockwise to decrease the speed.

When changing the direction of flow, turn the ccwcw switch to its central off position until the
pumphead rotor stops; then turn it to the required direction of rotation which will re-start the pump.

If returning from auto control to manual control, disconnect the process signal from the pump.

In the event of a power failure, the pump will automatically re-start when the power is
returned. This will occur under manual or remote control.

IP21

ccw position (

I

II) means the rotor

will rotate counter-clockwise at its
set speed.

cw position (II

I

) means the rotor

will rotate clockwise at its set speed

Centre position (I

I

I) means that

the pump rotor will be stopped

0