beautypg.com

Triplett 1401 User Manual

Page 3

background image

7-10-3 Connect the test leads to the two points of the circuit under test.

Press the

SELECT button once the continuity symbol appears

on the top of the LCD and the “

Ω” units appear on the lower right of

the display.

7-10-4 If there is continuity (resistance is lower than approximately 50Ω)

the built-in beeper will sound.

NOTE:

1. Do not attempt this measurement on energized circuits.

2. If the leads are reversed, not connected, or connected to a resistance

more than approximately 2000Ω, “1” (over range) should be displayed

7-11 Measuring Capacitance

7-11-1 Connect the black test lead to COM jack and red lead to VΩmA jack.

7-11-2 Set the rotary switch at the desired “Ω╟

” range position

7-11-3 Connect test leads across the capacitance to be measured.

7-11-4 Press the SELECT button three times, until the “n F” units appear on

the lower right of the display.

7-11-5 Read the capacitance from the LCD.

NOTE: Capacitors should be discharged before being tested.

7-12 Non Contact AC Voltage detection

7-12-1 Set the rotary switch at the desired “NCV” range position.

This activates the Non Contact AC Voltage (NCV) detection circuit,

and the NCV green LED will illuminate. The “

NCV” symbol will

appear on the display along with “EF”.

7-12-2 Position the meter so the top is parallel to and in contact with, or

close to the conductor. When the AC voltage ≥90V AC RMS, the

NCV red LED and green LED will flash and the buzzer will sound.

NOTE:

1. Even if the meter doesn’t indicate the presence of AC voltage, the

voltage may still be present. Do not rely on the non-contact voltage

detector to determine the presence of live voltage. Voltage detection may

be affected by socket design, insulation thickness, shielding, and other

AC voltage field interaction. The NCV detector does not detect DC

voltages.

2. If the meter test leads are connected and measuring AC voltage, the NCV

detector circuit will also indicate AC voltage presence (LEDs and buzzer

activation) if NCV mode is activated.

3. Proximity of florescent lights, dimmable lights, motors, and other large

AC field generating devices can trigger Non-Contact AC Voltage

detection function and invalidate the test.

8. Battery Replacement

8-1 If the battery voltage drops below proper operation range the “ “

symbol will appear on the LCD display and the battery should be

changed.

8-2 Before attempting to remove the rear cover to replace the battery, be

certain the test leads have been disconnected from the circuit to avoid

electrical shock hazard. Power off the meter and remove the test leads

from the banana jacks. To access the battery, remove the (2) screws in

the rear cover and lift the cover.

8-3 Replace the old battery with the same type battery (1.5V AA. R6

or LR6).

8-4 Replace the rear cover of the meter and reinstall the (2) screws.

9. Fuse replacement

9-1 To avoid electrical shock or injury, remove test leads from circuitry

before replacing the fuse.

9-2 Before attempting to remove the rear cover to replace the fuse, be

certain the test leads have been disconnected from the circuit. Power off

the meter and remove the test leads from the banana jacks. To

access the fuse holder, remove the (2) screws in the rear cover and lift

the cover.

9-3 Replace the old fuse only with the same type and rating: 6×30mm

800mA/500V fast fuse or 10A/500V fast fuse

9-4 Replace the rear cover and reinstall the (2) screws.

10. Maintenance

10-1 Before attempting to remove the battery cover or open the case, be

certain that test leads have been disconnected from energized circuitry

to avoid electric shock hazard.

10-2 Replace the test leads if the any wire is exposed, or if they become

damaged.

10-3 Do not use harsh chemicals or solvents to clean the meter.

10-4 Do not use the meter with the rear cover removed. Injury may result.

10-5 If the meter will not be used for more than 1 month, remove the battery

to avoid possible damage from battery leakage.

7-5 Measuring AC Current

7-5-1 Connect the black test lead to COM jack and red lead to VΩmA jack

for maximum 800mA current. For a maximum 10A current, move the

red lead to the

10A jack.

7-5-2 Set the rotary switch to the desired “A ” range position. Press the

SELECT button once or until “AC” appears on the top of the LCD and

the “

mA” or “A” units appear on the lower right of the display. You can

press the

RANGE button to switch between measurement ranges

manually, or leave the meter in “Auto range” mode.

7-5-3 Observing polarity, connect test leads in series with the load under

measurement.

7-5-4 Read the current on the LCD.

NOTE:

1. When the quantity to be measured is unknown, be sure that the meter is

in “Autorange” mode and consider whether to use the 800mA or 10A input

jack before connecting test leads to the circuit.

2. When only “1” or “-1” is displayed, it indicates “over range”. Either select

a higher measurement range or do not continue to measure.

3. Current above 800mA on the

VΩmA jack will blow the fuse. The 10A jack

is fused at 10A. Do not apply more than 10A and do not apply 10A for

longer than 10 seconds. Allow a 15 minute cooldown.

4. If fuse blows, remove meter from circuit under test. Replace fuse by

removing the rear cover (2 screws) to access fuse.

7-6 Measuring Resistance

7-6-1 Connect the black test lead to COM jack and red lead to VΩmA jack.

7-6-2 Set the rotary switch to the desired Ω╟

range position.

7-6-3 Connect the test leads across the resistance under measurement

7-6-4 Read the resistance on the LCD.

NOTE:

1. DO NOT APPLY VOLTAGE WHEN MEASURING RESISTANCE.

2. If “1” or “-1” is displayed, it indicates over range. This means that the test

leads are not making connection to the circuit, or the resistance exceeds

the measurement ability of the selected range. Select a higher range to

obtain a measurement.

3. When checking in-circuit resistance, be sure the circuit under test has

all power removed and that all capacitors have been fully discharged

.

7-7 Measuring Temperature

7-7-1 Set the rotary switch to the °C/°F range position.

7-7-2 The LCD will show the present temperature of the meter.

7-7-3 To use the K-type thermocouple probe, connect the black banana plug

of the probe into the

COM jack and the red banana plug to the VΩmA

jack.

7-7-4 The temperature of the thermocouple probe tip is displayed on the

LCD

7-7-5 The probe included with the meter can be used for measuring

temperature up to 482

°F. To accurately measure higher

temperatures, use a high temperature rated K-type probe.

NOTE:

1. Do not connect the thermocouple probe to a voltage source

7-8 Battery Testing

7-8-1 Connect the black test lead to COM jack and red lead to VΩmA jack.

7-8-2 Set the rotary switch to the proper “ ” or “ ” range position to

test either a 1.5V or 9V battery.

7-8-3 Connect test leads to the + and - terminals of the battery under test

7-8-4 The LCD displays the battery’s voltage at the rated test current.

7-9 Diode Testing

7-9-1 Connect the black test lead to COM jack and red lead to VΩmA jack.

7-9-2 Set the rotary switch to the “Ω╟

” position.

7-9-3 Connect the red lead to the anode of the diode to be tested, and the

black lead to the cathode. Press the SELECT button two times, until

the diode symbol appears on the top of the LCD and the “V” units

appear on the lower right of the display

7-9-4 The diode’s forward voltage drop is displayed on the LCD.

NOTE

:

1. The meter will display the approximate forward voltage drop of the diode.

2. If the leads are reversed or not connected, “1” (over range) should be

displayed

7-10 Continuity Testing

7-10-1 Connect the black test lead to COM jack and red lead to VΩmA jack.

7-10-2 Set the rotary switch to the “Ω╟

” position.

This manual is related to the following products: