9 TROUBLESHOOTING

Problem

Probable cause

Remedy

The instrument is
always detecting
product

Rest of product adhered on the
fork tines.

Clean softly the fork with a brush,
avoiding to press on the tines.

Power supply not correct.

Verify that the power supply
cables are well connected to the
terminal strip and that the voltage
between them is correct.

Fused fuse.

Change the fuse (250 mA T).

The detector vibrates
but does not change its
state

Damaged instrument.

Contact with technical service

The detector changes
its state without
presence of fluid

Interferences in the installation

Contact with technical service

The detector does not
vibrate and the LED
inside the cover is off