2 of the metering body, Cleaning, Technical characteristics – Tecfluid DP Series User Manual
Page 8: 6 cleaning, 7 technical characteristics

8
5.2
Of the metering body
The following faults may be appear:
Magnetic coupling deterioration,
Deterioration of the torsion or reading spring,
Wearing out of torsion shaft bearings,
Wearing out or deterioration of the lever and target disk.
If any of these breakdowns occurs, it is preferable to repair it in TECFLUID S.A.
workshops, as a new calibration will be normally necessary.
6 CLEANING
In line
Pass the normal cleaning agent through the installation.
In the workshop
It should be done with soft brushes. Never use sharp objects or metal brushes.
7 TECHNICAL
CHARACTERISTICS
Accuracy:
DP65: ± 2.5% full scale
DP500: ± 4% full scale
Rangeability: 10:1
Scale length: ~100
mm.
Scales: Direct according to fluid to be measured or in %.
Operation:
Vertical with upwards flow (BD)
Vertical with downwards flow (DAB)
Horizontal with right to left flow (DES)
Horizontal with left to right flow (ED)
Connections: DP65 Between flanges.
DP500 EN 1092-1, ANSI, JIS flanges.
Other flange standards on request.
Materials:
Body: Plastic coated steel, EN 1.4404 (AISI 316L)
Indicator housing: IP65 in coated aluminium (standard)
EN 1.4404, PP, others on demand
Working
pressure:
PN40 DN40
…
DN80
PN16 DN100
…
DN300
PN10 DN250 … DN500 (DP500)
Other pressure ratings available on request.
Process temperature:
- Body in plastic coated steel: -20ºC … +130ºC.
- Body in EN 1.4404 (AISI 316L), PTFE bearings: -20ºC … +150ºC.
- Body in EN 1.4404 (AISI 316L): bronze bearings: -20ºC … +300ºC.