Replacement parts, Returning the unit to ssi, Replacement parts returning the unit to ssi – Super Systems Simple Dew User Manual
Page 11

Simple Dew Digital Dew Point Analyzer Operations Manual
Super Systems Inc.
Page 11 of 18
12.1
After the 11.3% (Zero) calibration has been completed, remove the sensor from
the calibration salt and replace the cap.
13.0
Re-assemble the unit
13.1
After the calibration process has been completed, remove the sensor probe from
the gland and return it to the sensor-sampling chamber, taking care to position it
properly. The white mark on the sensor probe should face towards the inlet tubing.
If the white mark is not visible, then it should be placed so the sample flow directly
strikes the face of the mirror on the sensor tip. In other words, the mirror should
face the incoming gas stream.
13.2
Hand-tighten the black sensor gland to prevent air from leaking out of the
sampling chamber.
14.0
Make sure that all caps are replaced on the calibration salts, and return the instrument
to service.
Replacement Parts
Description
Part Number
Factory Calibration
65010
Calibration Kit (non-traceable)
31030
Calibration Kit (NIST traceable)
31425
Bowl Filter
37048
Bowl Filter Element
31027
Dew Point Sensor
31038
Sensor Sampling Block
20192
Sample Tube
20263
Sample Tube (With High-Temp Filter)
20264
Complete Units
Simple Dew Unit (with Power Supply)
13134
Simple Dew Unit (without Power Supply) 13136
Returning the Unit to SSi
This analyzer contains some components that may require periodic replacement based on the
amount of use that the unit experiences and the methods in which it is used. If service on the
unit is necessary, it should be sent back to Super Systems, Inc. in the original packaging for
repair. If the original packaging is not available, the analyzer should be surrounded by impact-
absorbing materials and placed in a box. It is the responsibility of the shipper to ensure that the
Simple Dew arrives at SSi undamaged.