Sensoray Model 118 Smart A/D™ Instruction Manual

36

Specifications

9.2 Sensor Specifications

Notes

1.

Measurement resolution and accuracy is specified for the default (versus
high-speed) measurement mode. Derate resolution and accuracy by 75%
when using the high-speed measurement mode.

2.

Channels configured for 4-20mA current loops return data values
representing a percentage of full-scale current (4mA= 0%,
20mA=100%). For example, a 12mA current would produce the data
value 50.0, indicating 50% of full-scale.

3.

This is the default sensor type after a board reset. It is provided for
back-compatibility with earlier Sensoray Smart A/D products. Use of
this type is not recommended for new designs, as higher resolution is
available on the 5V, 200uV resolution range.

4.

Declaring a channel’s sensor type as Disabled removes the channel from
the scan loop. Data values returned from disabled channels are
indeterminate and should be ignored.

5.

Gage specifications assume 4-wire bridge circuit, 120

minimum

impedance, 10V excitation applied by sensor channel.

6.

Use the following function to express gage measurement resolution in
load units:

For example, take the case of a pressure gage that is rated at 3mV/V at
100 PSI full scale. The Smart A/D would be able to resolve a load to:

7.

SDC is the Sensor Definition Code that is used by the SetSensorType
command to declare a channel’s sensor type. See Section 6.2.1 for
details.

Table 9: Sensor Specifications

Sensor Type

Measurement

SDC

7

(hex)

Range

Resolution

1

Accuracy

1

Thermocouple

B

0 to +1820°C

0.1°C

3.3°C

24

C

0 to +1820°C

0.1°C

2.1°C

23

E

-270 to +990°C

0.1°C

0.8°C

01

J

-210 to +760°C

0.1°C

0.6°C

1B

K

-270 to +1360°C

0.1°C

1.0°C

1C

N

-270 to +1347°C

0.1°C

0.9°C

22

T

-270 to +400°C

0.1°C

0.6°C

1D

S

0 to +1760°C

0.1°C

3.0°C

1E

R

0 to +1760°C

0.1°C

2.8°C

1F

Thermistor

Omega 44006 or 44031

-55°C to +145°C

0.01°C

0.05°C

1A

RTD

Cu 10

, 0.0367

/°C

0 to +119°C

0.1°C

0.6°C

2C

Pt 100

, 0.385

/°C

-200 to +800°C

0.05°C

0.2°C

18

Pt 100

, 0.392

/°C

-200 to +800°C

0.05°C

0.2°C

19

Ni 200

, 1.10

/°C

-58.9 to +151.6°C

0.02°C

0.07°C

28

Ni 1K

, 5.60

/°C

-50.0 to +129°C

0.025°C

0.08°C

29

Gage

5

-500 to +500 mV

5µV

6

30µV

0F

DC Voltage

-5 to +5V

3

500µV

600µV

00

-5 to +5V

200µV

600µV

15

-500 to +500mV

20µV

40µV

16

-100 to +100mV

5µV

30µV

17

Current Loop

2

4 to 20mA

0.01%

0.08%

11

Resistance

0 to 400

0.02

0.04

0A

0 to 3K

0.125

0.25

14

0 to 600K

31

130

20

Disabled

4

Not Applicable — sensor removed from scan loop

13

Resolution

RatedLoad

2

10

6

×

(

)

RatedVoltageOut

×

---------------------------------------------------------------------------

=

Resolution

100PSI

2

10

6

×

(

)

0.003V

×

----------------------------------------------

0.017PSI

=

=