Specifications, General specifications, 9 specifications – Sensoray 118 User Manual
Page 37: 1 general specifications

Sensoray Model 118 Smart A/D™ Instruction Manual
35
Specifications
9 Specifications
9.1 General Specifications
Table 8: General Specifications
Category
Parameter
Condition
Specification
Power
Quiescent
+5VDC, ±5%
100mA, typical.
+12VDC, ±5%
35mA, typical.
-12VDC, ±5%
35mA, typical.
Temperatur
e
Operating range
Guaranteed
accuracy
-25°C to +85°C
Sense
inputs
CMRR
CMV < 5V, f < 1KHz
80dB, minimum
CMV
Guaranteed
accuracy
±5.0V, maximum
Input impedance
SCC disabled
1000 M
Ω
, minimum
SCC enabled
1 M
Ω
, typical
Input protection
Continuous
±20V CMV, maximum
2 seconds, max.
±70V CMV, maximum
A/D
converter
Type
Integrating, 16-bit resolution.
Measurement
period,
per channel
60Hz reject mode
(default mode)
16.7 milliseconds, typical (default mode).
4.0 milliseconds, typical (high-speed
mode).
50Hz reject mode
20 milliseconds, typical (default mode).
4.8 milliseconds, typical (high-speed
mode).
Total time slot,
per channel
60Hz reject mode
(default mode)
22 milliseconds, maximum (default mode).
9 milliseconds, maximum (high-speed mode).
50Hz reject mode
25.3 milliseconds, maximum (default mode).
11 milliseconds, maximum (high-speed mode).
Excitation
source
Pulse width
Applied for duration of channel time slot.
Type
Strain/pressure
gage
10VDC, 40mA maximum.
400
Ω
/3K
Ω
ranges
1.2mADC.
600K
Ω
range
5VDC in series with 4K
Ω
.
Bus
interface
Type
DTB slave, A16, D8
Address
requirements
Range
One 256-byte address block starting on any
256-byte address boundary.
Address Modifiers
Enabled by AM Codes 29 (short nonpriveleged
access) and 2D (short supervisory access).
May be customized by changing one PLD on the
board.