Error Message

Cause and Course of Action

Snr

There has been an error with the sensor. This can occur

if the test subject does not wait at least 20 minutes after

smoking, eating or drinking before using the unit or if

the unit has not been used for an extended period of

time. Press the Power On/Off button once to turn off the

unit and repeat operation steps 1 – 5. If this message is

displayed four or more consecutive times, the unit may

need to be serviced.

rEP

The unit did not detect an even alcohol concentration in

the breath sample. This can occur if the user varies the

blowing force when performing a test or if the exhaust port

is obstructed. Repeat operation steps 1 – 5 to perform

another test.

H

The unit detected a very high alcohol concentration

(above 0.40% BAC).

Lo

Low Battery. Install two (2) new AA batteries.

Err

The sensor has sustained damage; please contact us for

further support.

RECALL LAST RESULT

Follow these directions to recall your last test result. From the main

screen (temperature display) press the Left Button one (1) time. Your

last test result will display on-screen.

SET TEMPERATURE DISPLAY FORMAT

Follow these steps to set the temperature display format to Fahrenheit

°F or Celsius °C. The default setting is Fahrenheit °F.
Step 1) From the main screen (temperature display) press the Right

Button for two (2) seconds until “USr” is displayed on-screen.

Step 2) Press the Power Button one (1) time.
Step 3) Press the Left Button to choose your desired format: