K-Patents PR-23 Troubleshooting Guide User Manual
Page 10

TROUBLESHOOTING GUIDE
FOR PR-23 ALL MODELS
10 (14)
January 9, 2015
K-PATENTS OY
Postal Address: P.O.Box 77, FI-01511 Vantaa, Finland
Street Address: Elannontie 5, FI-01510 Vantaa, Finland
Tel. int.+358 207 291 570
Fax int.+358 207 291 577
www.kpatents.com
Vat No. FI03035575
Business ID 0303557-5
Registered in Helsinki
increase. The visible changes in the curve and the temperature depend on viscosity, steam pressure and
temperature, as well as the
software version of the transmitter. If the wash is not reaching the prism, whole
wash system should be examined. There might be blockage in the lines, the wash pressure might be too low
or there might be condensate in the lines because lack of insulation.
Figure 9
Example of successful prism wash
Best way to diagnose and verify successful washing is to log data from the instrument with FC-11. The FC-
11 logs all the diagnostic values, as well as, all the measurement values. As it can be seen from the picture,
the temperature (red) increases and the CALC (blue) decreases with every wash, whereas the CONC
(green) stays stable all the time. This also reveals intermittent problems in the washing systems.
Figure 10
Data shows clearly if the wash is successful or not.