6.3. DIAGNOSTIC MESSAGES
For diagnostic messages in the Indicating transmitter display, see the PR-01-S manual. The following mes-
sages appear only in divert systems:
-
"In divert control operation" or "Removed from divert control", Section 2.6.
-
"Solids warning" and "Solids alarm", Section 5.4.
-
** Divert control fault **
This message indicates that there is no communication to the divert control unit, probably cable error, Figure
3.12, cable 2.
If the instrument is not applied in a divert control system, de-activate the divert function, see end of Section
5.2.
Further the message ** Relay unit fault ** may appear. This indicates that the Relay unit PR-7080 functions
are defined, but no Relay unit is connected. Normally a Relay unit is not used in a Divert system because the
wash contacts are in the Divert control unit. Clear the relay functions by the key sequence Calibrate/Relay unit
and select an "0 Not Defined" for all four relays. For using an optional relay unit, see PR-01-S Manual, sec-
tion 9.1.
6.4. DIAGNOSTIC LIGHTS
The Divert circuit board, Figure 3.12, contains the following diagnostic LEDS:
D3
Processor cycles: should flash
D4
Processor reads serial bus: should flash
D10
Serial bus A input: should flash
D11
On if 24 V DC supply from A is OK
D12
Serial bus B input: should flash
D13
On if 24 V DC supply from B is OK
D14
On if external 24 V DC is applied
D19
Wash relay A on
D20
Wash relay B on
D21
Solids warning relay on
D22
Solids alarm relay on
D23
Divert relay on = not DIVERT (Section 2.2)
D24
Horn
on
D25
Refractometer
difference
relay on = difference alarm
D26
Override
D27
Valve to bypass
D28
Valve to burner