Setting temperature parameters, Table 1. parameters and controls for rcb20-plus – Hoefer RCB20-PLUS User Manual
Page 28

•
p20
Parameter
controlled
Limits
Function
Key sequence to
enter “Set” mode
Temperature
Variation Limit
3 – 9 °C
This setting allows you to
determine the amount of
variation allowed in the water
bath temperature before
an alarm is triggered. For
example when set to 3 °C, the
temperature alarm will trigger
if the water bath temperature
becomes 3 °C above or
below the desired water bath
temperature.
SET + ▲
(simultaneously)
Upper Temperature
Limit
50 – 110 °C
Triggers an alarm when the
water bath temperature reaches
the setting.
SET + ▼
(simultaneously)
Set Water Bath
Temperature
-10 to
+110 °C
Sets the desired internal
or external water bath
temperature.
SET
Change Water Bath
Temperature
Change the water bath
temperature while in use.
ADJ
Setting Temperature Parameters
The following temperature parameters can be controlled with the RCB20-PLUS:
• Temperature Variation Limit
• Upper Temperature Limit
• Water Bath Temperature
Table 1. Parameters and Controls for RCB20-PLUS