Specifications, Safety recommendations, Specifications 4. safety recommendations – Hoefer PR250 User Manual
Page 8

Note: the PR250 is a Universal
Voltage instrument, 100–240 V
50/60 Hz.
3. Specifications
Shaking speed
50 – 250 rpm
Orbit
20 mm
Digital timer
1 min – 99 hour 59 min or
continuous run
Max. load capacity
2.5 kg
Power supply
50 Watts
Power input
24 V DC 2.0 A
Dimensions
28 × 27 × 11 cm
(L × W × H)
Weight
7.5 kg
Temperature range
4º– 45º C
Relative humidity range
≤ 70%
Relevant EC Directives and Harmonized Standards:
Low Voltage Directive 2006/95/EC
EMC Directive 2004/108/EC
EN 61010-1 Lab Equipment
EN 61010-2-020 Lab Equipment
EN 61326-1:2006 Lab Equipment, EMC
4. Safety Recommendations
• Plug the instrument into a properly grounded outlet.
• Always disconnect the power cord before servicing.
• If this equipment is used in a manner not specified
by the manufacturer, the protection provided by the
equipment may be impaired.
• Only accessories and parts approved by or supplied
by Hoefer, Inc. may be used for operating,
maintaining, and servicing this product.
•
p2
Note: There are no motor brushes
to replace on the PR250, if motor
stops please contact the Hoefer
service department.