Static temperature regulation, Active temperature regulation – Hoefer SUB Series User Manual
Page 12

•
p4
Static Temperature Regulation
1
Attach a short length of rubber hose to
each connector.
2
Incline the unit at an angle of approximately
45 degrees with the ports uppermost.
3
Use a funnel to fill the Cooling Platform with
deionized water containing 0.02% (w/v) sodium
azide (preservative to prevent bacterial and
algal growth).
4
When filled, keep the unit inclined and attach
clamps to each piece of rubber hose.
5
The unit can be cooled before an electrophoresis
run if required.
DO NOT FREEZE.
Active Temperature Regulation
1
Attach a short length of rubber hose to
each connector.
2
Attach one end of the rubber hose to the outlet
port of a circulating water bath/flow cooler and
the other end to the inlet port. Alternatively
attach one end of a rubber hose to a water
supply and allow the other rubber hose drain
to waste.
3
The maximum recommended water flow rate is
1 Liter/min. DO NOT exceed this figure.
4
If you are using a circulating water bath/flow
cooler, which exceeds this flow rate, you can
attach a T-connector in line. One branch of
the connector can return water to the bath
and the other can flow to the Cooling Platform
and incorporate a flow regulator such as an
adjustable tubing clamp. Measure and adjust the
flow rate before attaching the line to the gel unit.