 p6

Final casting steps

1

Pour the agarose solution (cooled to 50 °C) onto the 
running tray seated in the casting tray. Orient the 
comb assembly so that it is at the end of the tray 
opposite the direction of migration (typically at the 
cathode [–] end, which is marked by a black dot on 
the handle). Fit the comb assembly into the slots.

2

Allow a minimum of 30 min for the gel to set, then 
remove the comb carefully: partially lift and slightly 
tilt the comb at one end and slowly withdraw it from 
the gel. (Pulling the comb straight up creates a 
vacuum in the wells that may lift the gel out of  
the tray.)

3

Lift the running tray out of the casting tray and 
transfer it with the gel to the horizontal unit. Orient 
the running platform so that the sample will “run to 
red”. That is, place the sample wells at the cathode 
(–) end, which is indicated by a black dot. A notch on 
either side of the running tray centers the tray on the 
running platform.