Mca527 – GBS Elektronik MCA-166 Communication DLL User Manual
Page 16

MCA527
Name
MMCA_SET_FLAT_TOP_TIME
Description
The function sets the flattop time.
Declaration
ERROR_FLAG CALLBACK MMCA_SET_FLAT_TOP_TIME(unsigned short unFlatTopTime);
unFlatTopTime
Flattop time:
0 … 50 * 0.1 µs
Return value
See MCA_COMM function
Remarks
This function is only usable for MCA527.
Name
MMCA_SET_JITTER_CORRECTION
Description
The function sets the jitter correction.
Declaration
ERROR_FLAG CALLBACK MMCA_SET_JITTER_CORRECTION(unsigned short unOffOn);
unOffOn
0 turns jitter correction off and
≠
0 turns it on
Return value
See MCA_COMM function
Remarks
This function is only usable for MCA527.
Name
MMCA_SET_BASELINE_RESTORING
Description
The function sets the baseline restorer.
Declaration
ERROR_FLAG CALLBACK MMCA_SET_BASELINE_RESTORING(unsigned char uchOption);
uchOption
See CMD_SET_BASELINE_RESTORING
Return value
See MCA_COMM function
Remarks
This function is only usable for MCA527.
Name
MMCA_SET_PILE_UP_REJECTION
Description
The function sets the pile up rejection.
Declaration
ERROR_FLAG CALLBACK MMCA_SET_PILE_UP_REJECTION(unsigned short unOffOn);
unOffOn
0 turns PUR off and
≠
0 turns it on
Return value
See MCA_COMM function
Name
MMCA_SET_FAST
Description
The function sets the fast discriminator threshold.
Declaration
ERROR_FLAG CALLBACK MMCA_SET_FAST(unsigned short unFast);
unFast
0 ... 2499, default 400
Return value
See MCA_COMM function
Name
MMCA_SET_SLOW
Description
The function sets the slow discriminator threshold.
Declaration
ERROR_FLAG CALLBACK MMCA_SET_SLOW(unsigned short unSlow);
unSlow
0 ... 2499, default 400
Return value
See MCA_COMM function