Ecg performance testing, Square wave, Triangle wave – Fluke Biomedical PS410 User Manual
Page 25: Pulse wave, Sine wave

ECG Simulator
Simulating Functions
17
ECG Performance Testing
Square Wave
Code Display
Sets:
22
2 Hz
2.0 Hz square waveform.
23
0.125 Hz
0.125 Hz square waveform.
Triangle Wave
Code Display
Sets:
24 TRI 2 Hz
2.0 Hz triangle waveform.
Pulse Wave
Code Display
Sets:
25
30 BPM P
Pulse of 30 BPM, width of 60 ms.
26
60 BPM P
Pulse of 60 BPM, width of 60 ms.
27
120BPM P
Pulse of 120 BPM, width of 60 ms.
Sine Wave
The Simulator fixes amplitude at 1.0 mV for sine waves.
Code
Display Selects:
69
0.5 Hz SI
0.5 Hz sine wave.
70
5 Hz SI
5.0 Hz sine wave.
71
10 Hz SI
10.0 Hz sine wave.
72
40 Hz SI
40.0 Hz sine wave.
73
50 Hz SI
50.0 Hz sine wave.
74
60 Hz SI
60.0 Hz sine wave.
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)