ECG Simulator
Simulating Functions
13
Code Display
Selects:
20 NSR PED
Pediatric NSR with QRS width of 40
ms.
21 NSR ADLT Adult NSR with QRS width of 80
ms.
Arrhythmias: Premature Beats
Code Display
Selects:
28
PVC1 *
PVC with left ventricle focus. The
Simulator then assumes NSR at
80 BPM.
29
PVC1 EAR * Early PVC with left ventricle focus.
The Simulator then assumes NSR
at 80 BPM.
30
PVC1 ROT * R on T PVC with left ventricle
focus. The Simulator then
assumes NSR at 80 BPM.
34
MULTIFOC * Multifocal PVCs. The Simulator
then assumes NSR at 80 BPM.
Code
Display Selects:
36
PVC2 *
PVC with right ventricle focus. The
Simulator then assumes NSR at
80 BPM.
37
PVC2 EAR *
Early PVC with right ventricle
focus. The Simulator then
assumes NSR at 80 BPM.
38
PVC2 ROT *
R on T PVC with right ventricle
focus. The Simulator then
assumes NSR at 80 BPM.
* The simulation event occurs only once. To repeat the
event, enter the selection again.
Arrhythmias: Ventricular
Code
Display Selects:
31
PVCs (2) *
Pair of PVCs. The Simulator then
assumes NSR at 80 BPM.
32
RUN 5 *
Run of 5 PVCs. The Simulator then
assumes NSR at 80 BPM.