Airway pressure, High-flow port – Fluke Biomedical VT Mobile Getting Started User Manual
Page 38

VT MOBILE
Getting Started Manual
28
Airway Pressure
Maximum Applied
Pressure:
5 psi
Operating Pressure:
- 20 cmH
2
O to 120 cmH
2
O
Span Accuracy:
± 2 % of reading or ± 2.0
cmH
2
O, whichever is greater
Frequency
Response:
> 25 Hz or t
10-90
< 40 ms
Resolution:
0.1 cmH
2
O
Sample Rate:
100 Hz
Fittings:
Internally connected to flow
sensor pressure lines
High-Flow Port
Maximum Flow Rate
(absolute value):
200 lpm
Operating Flow Range:
± 150 lpm
Accuracy:
± 3 % of reading or ± 2 %
of range, whichever is
greater
Floor for Absolute
Accuracy:
25 lpm
Resolution:
0.01 lpm
Frequency Response:
> 25 Hz or t
10-90
< 40 ms
Sample Rate:
100 Hz
Dynamic Resistance:
< 2 cmH
2
O @ 60 lpm
Low-Flow Dropout:
2.5 lpm
Breath-Detect
Threshold:
4 lpm
Volume Range:
> ± 60 l
Tidal Volume Accuracy: ± 3 % or reading or ± 20
ml, whichever is greater
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)