Fluke Biomedical VT Mobile Getting Started User Manual
Page 32

VT MOBILE
Getting Started Manual
22
8.
Apply power to all elements of the test setup. For the
Analyzer, press
H
on, wait for the Zeroing screen,
and then press
G
. The Analyzer is ready to use
once the Tidal Volume screen appears.
9.
Press the access keys shown in Table 5 to display
the various ventilator parameters. Once you have
pressed the first key, you can continue pressing that
key to view other parameters or press
G
to view
all the parameters.
As described in Table 5, the Analyzer calculates 16
breath parameters in Local Mode. Note that all 16
parameters and other information can be viewed on one
screen in Remote mode when using the VT for Windows
software.
Figure 10 shows some characteristic ventilator parameter
screens.
Figure 11 shows additional oxygen, temperature, relative
humidity, and barometric pressure parameter screens.
Figure 12 shows additional airway pressure measurement
screens accessible during the ventilator check.
Table 5. Ventilator Parameters
Keys Item
Name
VT Tidal
Volume
MV Minute
Volume
PIF
Peak Inspiratory Flow
K
G
PEF
Peak Expiratory Flow
PIP
Peak Inspiratory Pressure
PEEP Positive End-Expiratory Pressure
MAP
Mean Airway Pressure
L
G
IPP
Inspiratory Pause Pressure
Rate Breath
Rate
Ti Inspiratory
Time
Te Expiratory
Time
M
G
I:E
Inspiratory to Expiratory Time Ratio
P
O2 Oxygen
Percentage
Æ P
Temp Temperature (gas under test)
Æ P
RH
Relative Humidity (gas under test)
S E
Baro Barometric
Pressure