Gas Flow Analyzer
Controls and Indicators
7
Table 4. Controls and Indicators
No. Name
Description/Use
A
Miniature RS-232 Serial Port
Connect the RS-232 cable here (for use with VT for Windows
software.)
B
Battery Eliminator Connector
Connect the battery eliminator here. Use the battery eliminator
whenever possible.
C
Battery Eliminator Indicator
Glows green when the battery eliminator is supplying power to the
Analyzer. Note that this power source cannot recharge the battery.
D
High Pressure Connector
Connect the High Pressure Adapter here. The Analyzer makes high
pressure measurements from -2 to 100 psi.
E
Flow and Low Pressure Connector
Insert one of three coded modular connectors here; the Analyzer
automatically detects the connector type. The Analyzer use the
High-Flow Sensor to measure up to
± 150 lpm, the Low-Flow
Sensor to measure up to
± 25 lpm, or the Low-Pressure Adapter to
measure from -20 to 120 cmH2O.
F
Display Screen
Shows measurement parameters and statistics (Min, Avg, Max),
waveforms, and setup selections. English only.
G
Oxygen Sensor Connector
Connect the Oxygen Sensor here. The Analyzer makes O2
measurements from 0 to 100%.
H
Temperature, RH Sensor Connector
Connect the optional Temperature and Relative Humidity Sensor
here.