Detailed specifications – Fluke Biomedical ESA609 User Manual
Page 42

ESA609
Users Manual
30
Detailed Specifications
Test Standard Selections ........................................ ANSI/AAMI ES-1/NFPA99, IEC 62353, and IEC 60601-1
Mains Voltage Measurement
Range ................................................................. 90.0 V ac rms to 264.0 V ac rms
Accuracy .............................................................
±
(2 % of reading + 0.2 V)
Earth Resistance
Modes ................................................................. Two wire
Test Current ........................................................ >200 mA dc
Range ................................................................. 0.000
Ω
to 20.000
Ω
Accuracy .............................................................
±
(1 % of reading + 0.010
Ω
)
Equipment Current
Range ................................................................. 0.0 A ac rms to 20.0 A ac rms
Accuracy .............................................................
±
(5 % of reading + (2 counts or 0.2 A, whichever is greater))
Duty cycle ........................................................... 15 A to 20 A, 5 min. on/5 min. off
10 A to 15 A, 7 min. on/3 min. off
0 A to 10 A, continuous
Leakage Current
Modes ................................................................. True-rms
Patient Load Selection ........................................ AAMI ES1-1993 Fig. 1
IEC 60601: Fig. 15
Crest factor .........................................................
≤
3
Ranges ............................................................... 0.0
μ
A to 1999.9
μ
A
Accuracy
DC to 1 kHz ....................................................
±
(1 % of reading + (1
μ
A, whichever is greater))
1 kHz to 100 kHz ............................................
±
(2.5 % of reading + (1
μ
A, whichever is greater))
100 kHz to 1 MHz ...........................................
±
(5 % of reading + (1
μ
A, whichever is greater))