Performing a current consumption test, Performing leakage current tests – Fluke Biomedical ESA609 User Manual
Page 28

ESA609
Users Manual
16
Performing a Current Consumption Test
To measure the current consumed by the DUT, press .
The Analyzer displays the current flowing through the
mains connections of the test receptacle.
Performing Leakage Current Tests
The Analyzer measures leakage current for a number of
different DUT configurations.
The leakage tests that are available depend on which
standard is selected. See the “Selecting the Test
Standard” section earlier in this manual to change the
standard the Analyzer is using.
Table 5 lists tests that have different names based on
which standard is selected.
Table 5. Test Names Based on Selected Standard
IEC60601 AAMI/NFPA
99
Protective-Earth
Resistance
Ground-Wire Resistance
Earth-Leakage Current
Ground-Wire Leakage
Current
Touch- or Enclosure-
Leakage Current
Chassis-Leakage Current
Press to access the leakage current main test shown in
Figure 9.
fis217.bmp
Figure 9. Leakage Current Test
Note
The display shown in Figure 9 is a leakage current
test when AAMI is the selected standard.
All leakage currents are displayed in as RMS only. The
initial result is displayed in the appropriate parameter
based on the standard selected.