Unpacking the analyzer – Fluke Biomedical ESA609 User Manual
Page 16

ESA609
Users Manual
4
•
Use only the mains power cord and
connector approved for the voltage and
plug configuration in your country and
rated for the Product.
•
Make sure the ground conductor in the
mains power cord is connected to a
protective earth ground. Disruption of the
protective earth could put voltage on the
chassis that could cause death.
•
Replace the mains power cord if the
insulation is damaged or if the insulation
shows signs of wear.
•
Connect the common test lead before the
live test lead and remove the live test lead
before the common test lead.
•
Keep fingers behind the finger guards on
the probes.
•
Do not use test leads if they are damaged.
Examine the test leads for damaged
insulation and measure a known voltage.
•
Do not use a current measurement as an
indication that a circuit is safe to touch. A
voltage measurement is necessary to know
if a circuit is hazardous.
Unpacking the Analyzer
Carefully unpack all items from the box and check that you
have the following items:
•
ESA609
•
Safety
Sheet
•
Users Manual CD
•
Carrying
case
•
Power
cord
•
ESA USA Accessory Kit (USA, Australia, and Israel
only)
•
ESA EUR Accessory Kit
•
Null Post Adapter