Zeroing the monitor and tester – Fluke Biomedical DPM1B User Manual
Page 19

Pneumatic Transducer Tester
Operation
Overview
9
Pneumatic
Transducer
Tester
BP Input
Filter
Monitor
2
1
ehf2.eps
Figure 2. DPM1B to Monitor Setup
3. When the system is to be filled with liquid, position Stopcock 1
as shown in Figure 2.
4. Attach a liquid-filled syringe to the top port of Stopcock 2.
5. Position the valve on Stopcock 2 so that the syringe can be used
to fill the system with liquids. (Turn the valve on the stopcock so
that the two opposing end arrows point straight up and down and
the middle arrow points toward the Tester, as depicted in
Figure 2.)
6. After filling the system, position both Stopcocks as shown in
Figure 2.
7. Remove syringe from top part of Stopcock 2.
Proceed to the section labeled Zeroing the Monitor and Tester.
Zeroing the Monitor and Tester
Properly zeroing the measurement system is important to accurately
determine transducer sensitivity.
1. Position the valve on Stopcock 1 as shown in Figure 4.
2. Press
the
ZERO or BALANCE control on the monitor.
3. Turn the Tester ON using the ON/OFF switch on the front
panel.