Fluke Biomedical 35080B User Manual
Page 67

Appendix
Procedure
C
C-3
Table C-2. Look-up-table for Linear Mammo Filter Pack (37351) measurements made on Rh
anode, Rh filtration selection of DMR mammography system. The equation used for
the correction is
Corrected kVp = (Uncorrected kVp x 1.2) - 12.5
Uncorrected
kVp
Corrected
kVp
Uncorrected
kVp
Corrected
kVp
Uncorrected
kVp
Corrected
kVp
30.5 24.1 33.9 28.2 37.3 32.3
30.6 24.2 34.0 28.3 37.4 32.4
30.7 24.3 34.1 28.4 37.5 32.5
30.8 24.5 34.2 28.5 37.6 32.6
30.9 24.6 34.3 28.7 37.7 32.7
31.0 24.7 34.4 28.8 37.8 32.9
31.1 24.8 34.5 28.9 37.9 33.0
31.2 24.9 34.6 29.0 38.0 33.1
31.3 25.1 34.7 29.1 38.1 33.2
31.4 25.2 34.8 29.3 38.2 33.3
31.5 25.3 34.9 29.4 38.3 33.5
31.6 25.4 35.0 29.5 38.4 33.6
31.7 25.5 35.1 29.6 38.5 33.7
31.8 25.7 35.2 29.7 38.6 33.8
31.9 25.8 35.3 29.9 38.7 33.9
32.0 25.9 35.4 30.0 38.8 34.1
32.1 26.0 35.5 30.1 38.9 34.2
32.2 26.1 35.6 30.2 39.0 34.3
32.3 26.3 35.7 30.3 39.1 34.4
32.4 26.4 35.8 30.5 39.2 34.5
32.5 26.5 35.9 30.6 39.3 34.7
32.6 26.6 36.0 30.7 39.4 34.8
32.7 26.7 36.1 30.8 39.5 34.9
32.8 26.9 36.2 30.9 39.6 35.0
32.9 27.0 36.3 31.1 39.7 35.1
33.0 27.1 36.4 31.2 39.8 35.3
33.1 27.2 36.5 31.3 39.9 35.4
33.2 27.3 36.6 31.4 40.0 35.5
33.3 27.5 36.7 31.5 40.1 35.6
33.4 27.6 36.8 31.7 40.2 35.7
33.5 27.7 36.9 31.8 40.3 35.9
33.6 27.8 37.0 31.9 40.4 36.0
33.7 27.9 37.1 32.0
33.8 28.1 37.2 32.1