1 introduction, 2 specifications, 3 theory of operation – Fluke Biomedical 35080B User Manual
Page 31

Linear Mammo Filter Pack Option (37351)
Introduction
6
6-1
Section 6
Linear Mammo Filter Pack Option (37351)
6.1 Introduction
The Linear Mammo Filter Pack (37351) provides a non-invasive way to make kVp measurements of
mammographic x-ray generators. This filter pack features nearly perfect linearity for precision
measurements between 22 and 40 kVp (for beryllium window, molybdenum anode mammographic
generators with molybdenum filtration).
6.2 Specifications
Intended Use
Beryllium window, molybdenum anode mammographic generators
Range
22.0 kVp to 40.0 kVp
Accuracy
± 1.0 kV exclusive of filtration effects
Sensitivity
0.01 volt per kV
Cal
0.200 V (20.0 kV)
Minimum Anode Current
50 mA
Target to kVp Divider Distance
12 in to 24 in (30.5 to 61 cm)
6.3 Theory of Operation
The Linear Mammo Filter Pack (37351) is similar in construction to the Wide Range linear diagnostic filter
pack except that the filter elements are much thinner. As expected in mammography, measurements are
more sensitive to variations in added filtration than in the diagnostic range. Thus, the Linear Mammo
Filter Pack (37351) is normally used in conjunction with the K-edge Mammo Filter Pack (37355). The K-
edge Mammo Filter Pack (37355) is used first to find a point between 27.5 kV and 29.5 kV very
accurately. Then the Linear Mammo Filter Pack (37351) measurement is made at the same point to
determine the necessary correction to apply to the Linear Mammo Filter Pack (37351) readings.