5 replaceable parts information – Fluke Biomedical 35080B User Manual
Page 54

35080B
Operators Manual
10-6
Figure 10-3.
Effects of RP103 Adjustment
10.5 Replaceable Parts Information
Fluke Biomedical maintains a complete inventory of all normal replaceable parts (see Table 10-1). To
place an order, or to obtain information concerning replaceable parts, contact the Fluke Biomedical at
440.498.2564 or by Fax at 440.542.3682.
Table 10-1.
Replaceable Parts List
Description
Part Number
Power battery, 9.0 V, alkaline
16-29
Wide Range Filter Pack
37617
Test Filter Pack (Optional)
32863
Linear Mammo Filter Pack (Optional)
37351
Cadmium K-edge Mammo Filter Pack (Optional)
37355
CT Filter Pack (Optional)
33551
Mobile Filter Pack Plus (Optional)
37946
Low Range Filter Pack (Optional)
38237
Carrying Case
37423
Battery Cover
37842
Rubber Feet
38471
LEGEND
kV
Waveform
Model 35080B Waveform Output Before
Adjustment of RP103
Increased Amount of Leading and
Trailing Edge in the Model 35080B
Waveform After Adjustment of RP103
(Turned Clockwise)