Attention, Tools – Flowserve LRGT16-1 Spector compact Conductivity Transmitter System User Manual
Page 6

Reducing length of installation and measuring length (control range)
The measuring tube and the electrode can be cut to suit the individual requirements of the
installation.
Length of installation A
[mm]
Measuring length L
[mm]
Possible reduction
of length R [mm]
215 200
315 300 50
415 400 50
515 500 50
615 600 50
815 800 100
1015 1000 100
1. Carefully determine the required reduction of length - measured from the bottom end of
the measuring tube - and mark this on the measuring tube.
2. Only slacken locking grub screw using a screwdriver size 1.3 for socket-head cap
screws.
3. Unscrew measuring tube and cut it to the length marked.
4. Unscrew spacer disc.
5. Mark the reduction of length - measured from the extreme end of the tip - on the
electrode tip.
6. Cut electrode tip to length. Cut back 10 - 12 mm of the insulation.
7. Screw back the spacer disc onto the electrode tip and fasten fingertight with a 12 mm
open end spanner.
8. Screw back measuring tube with a torque of 50 Nm and tighten locking grub screw.
Attention
!
The seating surfaces and threads on the vessel and mounting flanges must be
accurately machined.
!
Do not bend the electrode tip when cutting. Do not subject electrode to physical
shocks.
!
Do not lag the electrode body.
!
The conductivity transmitter system can be installed horizontally or with a vertical
inclination. The measuring surface must be permanently submerged.
Tools
!
41 mm open-end spanner
!
12 mm open-end spanner
!
Screwdriver size 1.3 for socket-head cap screws
!
Hacksaw
!
Flat file, cut 2
6