Display menu, Electrical connection – Bronkhorst MASS-VIEW Series Quick Start User Manual
Page 7
Settings
back
marker
Main
measure
fluidset
alarms
counter
settings
security
info
back
marker
enable
Marker
value
70
%
back
type
vol (n)
Measure
unit
ln/min
readout
percentage
back
fluid
air
Fluidset
capacity
20.00
ln/min
zero
back
reset
Counter
cancel
Settings
activate
min + max
Alarms
set min
65
%
set max
83
%
delay
3
sec
relais
pulse
mode
manual
reset
back
reset
Counter
mode
up to..
limit
1000000
unit
ln
defaults
sleep
on
delay
5
min
contrast
2
flipscrn
comm.
RS-232
back
Security
marker
disable
measure
disable
fluidset
disable
alarms
disable
counter
disable
settings
disable
back
model
mv-106
Info
serial
SN7654321
usertag
usertag
comm.
RS-232
version
V1.00
back
restore
new pass
cancel
Air
ln/min
Total: mln
3.717e4
Flow: ln/min
16.00
20.0
10.0
0.0
Display menu
MASS-FLOW ONLINE BV www.massflow-online.com
BHT20
10-03
6
Male
Female
1
1 TX-RS232 / D0 Modbus (A/A')
2
2 Analog output
3
3 Alarm high
4
4 0V power
5
5 Alarm low
6
6 RX-RS232 / D1 Modbus (B/B')
7
7
+
Us
8
8 0V sense
Shield
Output
: 0...5 Vdc
Supply voltage : +15...+24 Vdc
Note: 0V power (pin 4) and 0V sense (pin 8)
should be separately connected to the 0V
terminal at the power supply.
0Vdc +Us
power supply
Electrical connection