3 basic operation, 1 general, 2 analog operation – Bronkhorst IQ+FLOW (from 01-07-2013) User Manual
Page 12: Basic operation, General, Analog operation, 3basic operation

Bronkhorst High-Tech B.V.
IQ+FLOW
9.17.045
12
3
Basic operation
3.1
General
An IQ
+
FLOW® instrument must be powered with +15…+24 Vdc according to the applicable hook-up diagram, supplied with the
instrument. The instrument can be operated by means of:
Analog interface: 0…5 Vdc; 0…10 Vdc; 0…20 mA or 4…20 mA (single-channel versions only)
Digital RS232 interface (FLOW-BUS (Propar) protocol)
Digital RS485 interface (Modbus RTU, Modbus ASCII or FLOW-BUS protocols)
By default the instrument is set as specified. The table below lists the supported interfaces for single-channel and multi-channel
versions.
Analog interface
(section 3.2)
Digital RS232 interface
(section 3.4)
Digital RS485 interface
(section 3.5)
IQF/IQP
(single-
channel)
0…5 Vdc; 0…10 Vdc;
0…20 mA; 4…20 mA
(software selectable)
FLOW-BUS (Propar) protocol on 9600,
19200, 38400, 57600 or 115200 Baud
(software selectable)
Modbus RTU and Modbus ASCII protocols
on 9600, 19200, 38400, 56000, 57600 or
115200 Baud; FLOW-BUS protocol on
187500 or 400000 Baud (software
selectable)
IQM
(multi-
channel)
Not supported
FLOW-BUS (Propar) protocol on 38400 or
115200 Baud (select with rotary switch)
Modbus RTU protocol on 9600, 19200 or
38400 Baud (select with rotary switch)
3.2
Analog operation
Analog operation is only possible with single-channel IQ
+
FLOW® instruments. Multi-channel instruments can only be operated
digitally. At analog operation the following signals are available:
Measured value (analog output) at pin 2
Setpoint (analog input/setpoint) at pin 3
The valve output is no longer available as analog signal with IQ
+
FLOW®. The selected analog interface (0…5 Vdc; 0…10 Vdc; 0…20
mA or 4…20 mA) can be found in the model key of the instrument. Refer to Section 2.13.
When operating the instrument through the analog interface it is possible to connect the instrument simultaneously
to RS232 for reading/changing parameters (e.g. settings or fluid selection).
Hook-up
For analog operation either an RJ-45 loose-end cable or an RJ-45 to 9-pin sub-D converter may be used to connect the required
signals.
Refer to the “Hook-up diagram IQ
+
FLOW®” or use an
RJ-45 loose-end cable (7.03.419) to connect the required
signals.
When using a Bronkhorst readout unit use only an RJ-45
cable (7.03.236) in combination with the RJ-45 to 9-pin
sub-D converter (7.03.376). With these items the pin
configuration is unchanged.