What if there are problems, Technical specifications – Beurer BF 100 User Manual
Page 24

24
Observe the local regulations for material disposal.
Dispose of the unit in accordance with EC Di-
rective 2002/96/EC–WEEE (Waste Electrical
and Electronic Equipment).
If you have any questions, please contact the local au-
thorities responsible for waste disposal.
16. What if there are problems?
If the scale detects and error when measuring, the fol-
lowing is displayed.
Display
Cause
Remedy
ERROR:_
OVERLOAD
The maximum load
capacity of 150 kg
was exceeded.
Do not load the unit
with over 150 kg.
LOW_
BATTERY
The batteries are
almost empty.
Replace the batter-
ies (see page 16
and 23).
MEMORY_
FULL
30 measurements
have been saved.
Transfer the meas-
urements to a PC,
see page 23.
ERROR:_
CONTACT
The electrical re-
sistance between
the electrodes and
the sole of your
foot is too high
(e.g. if you have
tough calluses).
Repeat the meas-
urement barefoot. If
necessary, moisten
the soles of your feet
slightly. If necessary,
remove the calluses
from the soles of
your feet.
ERROR:_
DATA
The BIA measure-
ment was success-
ful but the fat per-
centage is outside
the measurable
range (less than
5% or greater than
50%).
Repeat the measure-
ment barefoot or, if
necessary, moisten
the soles of your feet
slightly. If necessary,
change the level of
activity that you have
set.
No display The batteries in
the control unit are
completely empty.
Replace the batter-
ies (see page 16).
The batteries in the
control unit are not
inserted correctly.
Check to ensure that
the polarity is correct
(see page 16).
No measurement possible
Possible errors
Remedy
The control unit
was not activated
before you stepped
on the scale.
Switch on the control unit first
before the measurement. If nec-
essary, wait 30 seconds, then re-
peat the measurement.
Possible errors
Remedy
Cordless connec-
tion between scale
and control unit is
interrupted.
Reinsert the batteries for the
scale and the control unit. Main-
tain a larger distance from the
following electromagnetic sourc-
es: high-voltage pylons, televi-
sion or wireless computer sys-
tems, fitness equipment, mobile
phones, weather stations.
The control unit is
too far away from
the scale.
Reduce the distance between the
scale and control unit; maximum
distance is 2 metres.
The batteries in the
scale are flat or in-
correctly inserted.
Change the batteries (see page
16) and check that they are in-
serted the correct way round.
Incorrect weight measurement
Possible errors
Remedy
The scale is not
zeroed correctly.
Before taking a measurement,
activate the scale briefly with
your foot until the
operating
status indicator lights up.
Wait approximately 5 seconds
and repeat the measurement.
For your information: because
you would normally step onto
the scale immediately, the scale
saves the zero point when the
scale is switched off (when the
operating status indicator
goes out).
Scale is on
carpeting.
Place the scale on a level and
solid surface.
Stand with your feet as far apart
as possible without tipping the
scales.
17. Technical specifications
Dimensions of the scale
330 x 330 x 41 mm
Scale batteries
3 x 1.5 V type AA
Control unit batteries
3 x 1.5 V type AA
Cordless transmission
433 MHz
Measuring range
5 kg to 150 kg
11 lb to 330,6 lb
0.79 St. to 23.61 St.
Weight, bone mass unit
100 g / 0.2 lb / 0.2 lb
Unit for body fat, water
and muscle
0.1 %
BMR, AMR unit
1 Kcal
Subject to technical modifications.
The scale conforms to EC Directive 2004/108/EC +
supplements.