Beurer GL 50 evo (mgdl) User Manual
Page 34

34
Beurer GL50 evo mg/dL
9 t
echnical
specifications
Dimensions (L x W x H)
123 x 29 x 16 mm
Weight
36 g
Power supply
Lithium-ion rechargeable battery, 160 mAh
Battery life
150 measurements with fully-charged rechargeable battery
Measured value memory
480 measured values with date/time
Data retained when batteries are changed
Average values
for 7, 14, 30, 90 days
Automatic switch-off
Two minutes after last actuation
Storage/
transport temperature
Temperature: +2°C – +30°C
Relative humidity: < 90%
Operating ranges
Temperature: +10°C – +40°C
Relative humidity: < 90% non-condensing
Measuring range, glucose
Glucose: 20 – 630 mg/dL
Blood sample
capillary whole blood, venous whole blood
Required volume of blood
0.6 microlitres
Blood glucose measuring duration Approx. 5 seconds
Calibration
Plasma
Test procedure
Amperometric bio sensor
Usage
Suitable for self-testing
System function test
Each time device is switched on
EMC
This device complies with the European standard EN 61326 and is subject to specific precautions
with regard to electromagnetic compatibility. Please note that portable and mobile HF communica-
tion systems may interfere with this unit. For more details, please contact Customer Services at the
address indicated.
Test strip functionality
Test strips enable a quantitative measurement of the glucose level in fresh whole blood. When the gap
for taking blood comes into contact with a drop of blood, it is automatically filled by simple capillary
action. The blood is sucked into the absorbing gap on the test strip and the blood glucose monitor
measures the blood glucose level in the blood.
The test is based on the measurement of an electric current that is generated by the chemical reaction
of the glucose with the enzyme glucose dehydrogenase (Aspergillus oryzae) on the strip.
During the reaction, a mediator transports electrons through the electrode surface and so gener-
ates a current.
The blood glucose monitor analyses this current. The current flow is proportional to the glucose con-
tent in the blood sample. The results are shown on the blood glucose monitor display.