Beurer GL 50 evo (mgdl) User Manual
Page 32

32
Beurer GL50 evo mg/dL
Note
The blood glucose monitor is made of precision components. The accuracy of the measurements
and service life of the device depend on its careful handling:
• Protect the device from impacts and do not drop it.
• Protect the device from damaging factors such as moisture, dirt, dust, blood, control solution or
water, marked temperature fluctuations, direct sunlight and extreme cold.
• Do not use the device in the vicinity of strong electromagnetic fields, radio systems or mobile
telephones.
8 W
hat
if
there
are
problems
?
Display messages on batteries and blood glucose measurement
No.
Cause
Solution
Empty
battery
Batteries discharged.
Charge the blood glucose monitor via the USB
interface.
High
temperature
Very high glucose level
(> 630mg/dL)
Temperature of the measuring environment,
blood glucose monitor or test strip above the
permitted range.
Repeat the test using a new test strip as
soon as the measuring environment, blood
glucose monitor and test strips have reached
room temperature (+20°C to +26°C).
Low
temperature
Very low glucose level
(< 20mg/dL)
Low temperature Temperature of the measur-
ing environment, blood glucose monitor or test
strip below the permitted range.
Repeat the test using a new test strip as soon
as the measuring environment, blood gluco-
se monitor and test strips have reached room
temperature (+20°C to +26°C).
Strip
error
Used or contaminated test strip
inserted.
Insert unused test strip that has not expired.
Repeat the blood glucose measurement.
Error 1
System error
Remove batteries, reinsert batteries.
Contact our customer service if the problem
persists.
Error 2
Insufficient amount of blood on
the test strip.
Repeat the measurement using a new test
strip.