Specifications – Beurer BM 70 User Manual
Page 23

23
• the cuff tube is not properly inserted (
appears in the dis-
play),
• inflation takes longer than 15 seconds (
appears in the dis-
play),
• the inflation pressure is higher than 300 mmHg (
appears
in the display),
• an error occurs when storing the measured values (
ap-
pears in the display),
• exceeding measurement range ( appears in the display).
In the about cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do not
move or talk. Put the batteries back in if necessary, or else re-
place them.
9. Specifications
Model no.
BM 70
Measurement
method
Oscillometric, non-invasive blood
pressure measurement on the upper
arm
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 30 – 260 mmHg,
diastolic 30 – 260 mmHg,
Pulse 40 –199 beats/minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement inac-
curacy
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 156 mm x W 117 mm x H 80 mm
Weight
Approx. 535 g (without batteries)
Cuff size
24 to 36 cm
Permissible operating
conditions
+10 °C to +40 °C, 40 – 85 % relative air
humidity (non-condensing)
Permissible storage
conditions
-10 °C to +60 °C, 10 – 90 % relative
air humidity, 800 – 1050 hPa ambient
pressure
Power supply
4 x 1,5 V
AA batteries
Battery life
For approx. 300 measurements,
depending on the blood pressure level
and/or pump pressure
Accessories
Instruction for use, 4 x 1.5 V AA batter-
ies, storage pouch
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied
part
These specification are subject to change without notice for
purpose of improvement.