Electromagnetic compatibility information – Beurer BC 16 User Manual
Page 86

86
Electromagnetic Compatibility Information
Table 1
For all ME EQuIPMEnT and ME sYsTEMs
Guidance and manufacture’s declaration – electromagnetic emissions
The BC 16 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BC 16 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The BC 16 uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The BC 16 is suitable for use in all establishments, including domestic establish
ments and those directly connected to the public lowvoltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions IEC 6100032
Not applicable
Voltage fluctuations/flicker emissions
IEC 6100033
Not applicable
Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
The BC 16 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 16 should assure that it
is used in such an environment.
Immunity test
IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 6100042
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30 %.
Electrical fast transient/burst
IEC 6100044
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for power
supply lines
N/A
Mains power quality should be that of a typical commercial or hos
pital environment.
Power frequency (50/60 Hz)
magnetic field IEC 6100048 3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic
of a typical location in a typical commercial or hospital environment.
noTE: U
T
is the a.c. mains voltage prior to application of the test level.