beautypg.com

Monroe Electronics Charge Plate - Ionizer Performance Analyzer - model 287A User Manual

Page 3

background image

Section 1

SPECIFICATIONS:

Charger:

1100 volts, selectable polarity

Fieldmeter:

Accuracy

5% of reading, 2% typical

Zero Drift

<

4V in 90s, 2V typical

Trip points:

Fixed 1000V and 100V

Charge plate:

1.7”x4” (

10.6 inches periphery)

Capacitance

25pF

Self-discharge

<10% of full scale in 200s

Display:

2x16 LCD, >5 updates per second

Timer

200.0s maximum, 0.1s resolution

Fieldmeter

1250V, 1V resolution

Sensors:

Humidity

5% typ. from 10% to 90%RH @25ºC

Temperature

2ºC typical

Oper. temp.:

25ºC,

10ºC

Battery:

9 volt NEDA #1604A or equivalent alkaline (>40 hour life or >1300
charge cycles). Longer life may be achieved by using 9-volt lithium.

Dimensions:

8.1x4x1.9 inches (206x102x48 mm)

Weight:

1lb, 6 oz. (0.63 kg)

Accessories included:

Carrying case

Banana patch cord

Alligator clip with boot

Battery

This manual and a manual supplement

What this instrument does —

Model 287a IPA meets and far exceeds the basic requirements of ESD Association
(Draft) Standard Practice 3.3 for the Protection of Electrostatic Discharge
Susceptible Items – Periodic Verification of Air Ionizers
.

Model 287a IPA performs manual or automatic decay and balance tests for periodic
verification of ionization equipment and stores the results and averaged decay times for
up to ten manual tests and up to ten complete automatic test sequences.

Provides real-time measurement and display of temperature and relative humidity.

Performs self-tests.

Battery powered and portable, sets up in seconds.