CHAPTER 9 FAULT FINDING
The following table outlines some possible fault conditions.
SYMPTOM FAULT
SUGGESTED
REMEDY
Failure to switch 
ON 
Battery not 
charged. 
Does the red 
charger LED 
illuminate? 
- Check that the battery is secured.
- Clean the battery connectors on the TUFF 
pump. 
- Clean the battery connectors on the charger.
- Push the TUFF pump firmly into the charger.
- Charger fault - return for repair. 
- Keypad fault - return for repair. 
Software locked 
up? 
Disconnect the battery for a minimum of 20 
minutes and then reconnect the battery. If this 
fails, return for repair. 
Pump runs fast
Control error
Damaged flow pressure sensor - return for 
repair. 
Water ingress - clean or return for repair. 
Tubes to sensor squashed or damaged - 
return for repair. 
Electrical fault - return for repair. 
Calibration error - perform fundamental 
calibration in Configuration Mode. 
Poor control 
response, 
Non-repeatable 
flow settings 
Leakage
Contamination or damage to valves 
- return for repair. 
Check all connections. 
If internal leakage is suspected - return for 
repair. 
Not getting 
expected run time 
from pump 
Higher back
pressure from
filter media
Ensure Tuff is fully charged. Consult filter back 
pressure tables to calculate correct run time. 
Often by selecting coarser filter material will 
offer considerably longer run times. 