beautypg.com

Equipment specification – Casella CEL CEL24X User Manual

Page 17

background image

12

.

EQUIPMENT SPECIFICATION

SPECIFICATION

CHARACTERISTICS

STANDARDS

IEC 61672-1 2002-5 (Electro-Acoustics – Sound Level
Meters) Group ‘X’ instruments, Performance Class 2.

IEC

60651: 1979 Type 2 .ANSI S1.4 type 2A Specification for
Sound Level Meters.

CEL-244/246 Models: IEC 60804: 2000 Type 2, ANSI
S1.43: 1997 (R2007) Type 2

RANGE:

Display range: 30 -130dB(A) RMS, available in 2 ranges,
30 -100dB and 60 -130dB. Linear operating range 10dB
above noise floor.

RMS FREQUENCY
WEIGHTINGS

A and C filter weightings, satisfying IEC 61672-1: 2002
Class 2, ANSI S1.4 Type 2A.

NOISE F LOOR

Total noise floor typically <33dB(A).

FREQUENCY RESPONSE

Overall frequency response as per IEC 61672-1: 2002
Class 2, ANSI S1.4 Type 2A.

TIME WEIGHTINGS

Fast, Slow and Impulsive according to IEC 61672-1:
2002, ANSI S1.4 Type 2A.

REFERENCE CONDITIONS

20oC air temperature, 65% relative humidity, 101.325kPa
atmospheric pressure. Nominal reference level = 114.0dB
at 1kHz. Free field perpendicular incidence.

OPERATING ENVIRONMENTAL CONDITIONS

Humidity

5 to 90% RH in the absence of condensation.

Temperature Range

0 – 40

o

C.

Pressure

65 to 108kPa.

Effects of Humidity

Less than ±0.5dB over the range 30 to 90% relative

humidity (non-condensing), relative to the value at

reference conditions.

STORAGE ENVIRONMENTAL CONDITIONS

Humidity

0 to 90% RH in the absence of condensation.

Temperature Range

-20 to +60oC.

Pressure

65 to 108kPa.

MICROPHONE

Internal electret capsule 10mV/Pa nom +/-3dB within ½”
preamplifier.

CALIBRATION

Auto calibration by application of 1kHz @ 114 or 94dB
calibrator (CEL-110/2 or CEL-110/1). ±1dB calibration
range.

POWER SUPPLY

External DC

5VDC (via US B 5 pin mini B).

Batteries

3x AA, Rechargeable NiMH or Alkaline cells.

17