American Weigh Scales (AWS) miniONYX User Manual
Page 6
miniONYX Manual
switch on the side of the scale to the CAL position.
The scale will turn ON and go to zero.
2. Hold the MODE key until the display shows “CAL”,
then release.
3. Tap the MODE key again. The display will flash
“CAL” followed by “500.0g”.
4. Place a 500g calibration weight on the center of the
platform. The display will flash “PASS”. Calibration
is complete!
5. You can now use the lock switch on the side to re-
turn the scale to normal (unlocked) weighing mode.
X. Troubleshooting
•
Problem: Scale will not turn on
•
Solution: Check the batteries
•
Problem: Displayed weight fluctuates randomly
•
Solution: Make sure your workspace is stable and
free of any air currents or vibration
•
Problem: Displayed weight is inaccurate
•
Solution: Perform a calibration adjustment
XI. Error Codes
1. OUTZ - Zero range has shifted. Perform calibration
adjustment. If the problem persists, the load cell may
be permanently damaged due to overloading.
2. 88888 - Maximum capacity exceeded. Remove the
extra weight to avoid damaging the load cell.
3. L - Batteries are low. Replace the batteries.