of an empty container or vessel for net weight determination.

NOTE: The weight of the container will subtract from the

available capacity.

1. Place the empty container or vessel on the weighing

platform.

2. Touch the

ON/T key once. The display will then reset

to zero. If the container is removed, it’s weight will

be displayed as a negative number.

3. You may now fill the container with the quantity to

be weighed. The display will show the net weight of

the contents.

4. To return to normal weight display (gross weight),

remove any items from the weighing platform and

touch the

ON/T key again to re-zero the scale.

VIII. Calibration
Calibration and adjustment are performed at the factory.

However, the weighing range can shift slightly during

shipping. Also, due to changes in local acceleration of gravity

around the world, adjustment is recommended when moving

the scale from one location to another. To check calibration:

1. Place the required calibration weight on the scale

and note the displayed value.

2. If the displayed value is not within acceptable

tolerance (see chart below), adjustment is required.

Model

Acceptable

Tolerance (±)

Calibration

Weight

Recommended

Accuracy Class

OXM-600

0.2g

500g

OIML M2

IX. Adjustment

1. To unlock calibration adjustment mode, set the lock