KG-Series Manual
1. Prepare the calibration weight (see pg. 4 for required calibration weights).
2. Clean the weighing pan and press TARE to re-zero the balance.
3. When the balance is stablized, press and hold the CAL button for 5 seconds.
4. The display will show “-CAL-”, then the required calibration weight will flash on the display.
5. Place the calibration weight onto the center of the platform and wait for the display to stop flashing.
6. You may now remove the weight. The scale will return to normal weighing mode
3.6. Linear Calibration
Linear calibration adjusts the balance at two or more points, rather than just one. To enter linear calibration,
follow these steps:
1. Prepare the calibration weights (see pg. 4 for required calibration weights).
2. Clean the weighing pan and press TARE to re-zero the balance.
3. When the balance is stablized, press and hold the CAL button for 5 seconds.
4. The display will show “-CAL-”, then flash the first required weight for span calibration.
5. At this point, press and hold the MODE button for 5 seconds. The display will flash the first required
calibration weight for linear calibration.
6. Place the required weight onto the center of the platform.
7. When the display stops flashing, you may remove the weight.
8. Once the weight is removed, the next required weight for linear calibration will flash.
9. Repeat steps 6 and 7 until the display returns to zero when the weight is removed - indicating that linear
calibration is complete.