Service Manual
3-8
Table 3-2. DC Voltage Test
Step
DC Input Voltage
Display Reading
1
+2.7V
2.691 to 2.709V DC
2
+27V
26.91 to 27.09V DC
3
+270V
269.1 to 270.9V DC
4
+1000V
995 to 1005V DC
5*
+300 mV
299.0 to 301.0V DC
*300 mV function only
3-8. AC Voltage Test
Warning
Connect the ground/common/low side of the AC
calibrator to COM on the UUT.
1.
Set the UUT function switch to VAC, and connect the AC Voltage
Calibrator to the
z and COM input terminals.
2.
Set the AC Voltage Calibrator for the output given in Table 3-3,
verify that the UUT display reading is within the limits shown.
Note
When the input is open in the VAC function, it is normal for
the meter to read some counts on the display. This is due to
ac pickup in the ac amplifier when the ac amplifier when the
ac amplifier is unterminated.
Table 3-3. AC Voltage Test
Input
Step
Voltage
Frequency
Display Reading
1
2.7V
100 Hz
2.644 to 2.756V AC
2
2.7V
500 Hz
2.644 to 2.756V AC
3
1000V
100 Hz
978 to 1022V AC
4
1000V
1000 Hz
978 to 1022V AC
3-9. Resistance
Test
1.
Select the ohms function on the UUT.
2.
Connect the Ohms Calibrator or Decade Resistor to the
z and
COM input terminals of the UUT.