True-rms Multimeters
General
Specifications
19
General Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18
°C to 28 °C, with
relative humidity at 0 % to 90 %.
Extended specifications are available at www.Fluke.com.
Maximum voltage between any
terminal and earth ground ...................................... 600 V
Surge Protection ..................................................... 6 kV peak per IEC 61010-1 600V CAT III,
Pollution Degree 2
W Fuse for A input (115 & 117 only): .................... 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293)
Display ..................................................................... Digital: 6,000 counts, updates 4/sec
Bar Graph: 33 segments, updates 32/sec
Temperature............................................................. Operating: -10 °C to + 50 °C
Storage: -40 °C to + 60 °C
Temperature Coefficient ......................................... 0.1 x (specified accuracy)/
°C (<18 °C or >28 °C)
Operating Altitude ................................................... 2,000 meters
Battery
....................................................................... 9 Volt Alkaline, NEDA 1604A / IEC 6LR61
Battery Life............................................................... Alkaline: 400 hours typical, without backlight
Safety Compliances ................................................ Complies with ANSI/ISA 82.02.01 (61010-1) 2004,
CAN/CSA-C22.2 No 61010-1-04, UL 6101B (2003)
and IEC/EN 61010-1 2
nd
Edition for measurement
Category III, 600 V, Pollution Degree 2, EMC
EN61326-1