beautypg.com

Spectrum Controls 2080sc-IF4u User Manual

Page 9

background image

Micro800™ 4 Ch Universal Analog Input Module

16 

Publication 0100189-02 Rev. A

Environmental Tests

Industry Standards

Test Level Limits

of initial contact)

2kV Air Discharge (to points of
initial contact)

Radiated RF immunity

(Performance Criteria A)

IEC 61000-4-3: Level 3

Rockwell Document QTP#X0327

10 V/M with 1 kHz sine-wave
80%AM from 80…2000 MHz

10 V/M with 200 Hz sine-wave
50% Pulse 100%AM @900 MHz

10 V/M with 200 Hz sine-wave
50% Pulse 100%AM @1890 MHz

1 V/M with 1 kHz sine-wave
80%AM from 2000…2700 MHz
(3 V/M goal)

EFT/B immunity

(Performance Criteria B)

IEC 61000-4-4*

Rockwell Document QTP#X0327

Signal Ports:

± 3 kV @ 5 kHz for 5 minutes,
Criteria B (Marine?)
± 2 kV @ 5 kHz for 5 minutes,
Criteria A (Marine?)

± 2 kV @ 5 kHz for 5 minutes,
Criteria B (standard)

Power Ports:

± 2 kV @ 5 kHz for 5 minutes,
Criteria A (Marine?)

± 2 kV @ 5 kHz for 5 minutes,
Criteria B (standard)

Surge transient immunity

(Performance Criteria B)

IEC 61000-4-5

Rockwell Document QTP#X0327

Signal Ports:

± 2 kV line-earth {CM}@ 2Ω on
shielded ports

Power Ports

± 2kV CM @ 12Ω
± 1kV DM @ 2Ω

Conducted RF immunity

(Performance Criteria A)

IEC 61000-4-6

Rockwell Document QTP#X0327

Not required for initial release.
Later will be tested to 10V rms with
1 kHz sine wave 80%AM from 150
kHz…80 MHz on signal and
power ports

Magnetic Field

(Performance Criteria A)

IEC 61000-4-8

Rockwell Document QTP#X0327

30Arms/m

AC Mains Voltage Dips,
Interruptions and Variations

IEC 61000-4-11

Rockwell Document QTP#X0327

Follow the 61000-4-11.

Micro800™ 4 Ch Universal Analog Input Module

Publication 0100189-02 Rev. A

Table 2 (Data Format)

Input Type

Input Value

Condition

EU x1

EU x10

E Thermocouple

1000.00 deg C High Range

10000

1000

-270.00 deg C

Low Range

-2700

-270

J Thermocouple

1200.00 deg C High Range

12000

1200

-210.00 deg C

Low Range

-2100

-210

K Thermocouple

1370.00 deg C High Range

13700

1370

-270.00 deg C

Low Range

-2700

-270

T Thermocouple

400.00 deg C

High Range

4000

400

-270.00 deg C

Low Range

-2700

-270

100 Ohm Pt 0.385

850.00 deg C

High Range

8500

850

-200.00 deg C

Low Range

-2000

-200

1000 Ohm Pt 0.385 850.00 deg C High Range

8500

850

-200.00 deg C

Low Range

-2000

-200

100 Ohm Pt 0.392

630.00 deg C

High Range

6300

630

-200.00 deg C

Low Range

-2000

-200

1000 Ohm Pt 0.392 630.00 deg C High Range

6300

630

-200.00 deg C

Low Range

-2000

-200

0..3000 Ohms

3000.00 ohms

High Range

30000

3000

0.00 ohms

Low Range

0

0

+/-50mV

50.00 mV dc

High Range

5000

500

-50.00 mV dc

Low Range

-5000

-500

+-100mV

100.00 mV dc

High Range

10000

1000

-100.00 mV dc

Low Range

-10000

-1000

0..5V

5.00 V dc

High Range

5000

500

0.00 V dc

Low Range

0

0

+/-10V

10.00 V dc

High Range

10000

1000

-10.00 V dc

Low Range

-10000

-1000

0..10V

10.00 V dc

High Range

10000

1000

0.00 V dc

Low Range

0

0

4..20mA

20.00 mA

High Range

20000

2000

4.00 mA

Low Range

4000

400

0..20mA

20.00 mA

High Range

20000

2000

0.00 mA

Low Range

0

0

CJC

85.00 deg C

High Range

850

85

-25.00 deg C

Low Range

-250

-25