beautypg.com

Cartridge loading adjustments – Simaudio 810LP Phono Preamplifier User Manual

Page 9

background image


810LP Reference Dual-Mono Phono Preamplifier

____________________________________________________________________________________

9

Cartridge Loading Adjustments

Prior to making any of these adjustments, always

disconnect the AC power cord and all interconnects

from MOON 810LP. We strongly recommend using

the

pen shaped

plastic hand tool

that we have

included with your 810LP as it was specifically

designed for this purpose. Using another tool made

from a material other than plastic may damage the

DIP switches and/or scratch the painted surface on

the bottom of the 810LP’. Finally, to achieve the

best possible sonic performance, it is absolutely

necessary that all settings be identical for both the

left and the right channels.

In the following examples,

the color

white always indicates the

position - left (‘ON’) or right

(‘OFF’) of the DIP switch.

In the

example to the left, the DIP switch is in the left

position (‘ON’).

This is to be consistent with the

actual white color of these switches.

Capacitance Loading:

There are sixteen (16) unique capacitance loading

settings available using DIP switches 7 through 10,

located at the top of the upper DIP switch bank. The

range of available settings begins at 0pF and ends at

1120pF:

Figure 2: Left and right channel DIP switches for capacitance loading adjustments

The factory default is 100 pF, whereby DIP switches

10, 9, and 8 are all in the right (‘OFF’) position and

DIP switch 7 is in the left position (‘ON’) as seen in

the example to the right. This setting

would be applicable for most moving

magnet cartridges.


When you are using a moving coil cartridge, you will

most likely want to set the capacitance loading to

0pF. This is done by positioning all four DIP switches

(7, 8, 9, & 10) to the right position (‘OFF’)

as seen in the example to the right.

A comprehensive diagram of the DIP switch

combinations for all 16 available capacitance loading

settings can be seen on the bottom panel,

immediately to the right of the switches. Finally, the

values written beside the DIP switch number
represent the capacitance load if only that switch is

in the left position (‘ON’). For example, when only

switch 10 is in the left position (‘ON’), the load value

will be 470pF; when only switch 9 is in the left

position (‘ON’), the load value will be 330pF; and

finally when only switch 8 is in the left position
(‘ON’), the load value will be 220pF.