High temperature module - htm – Parr Instrument HTM User Manual
Page 4
High Temperature Module - HTM
Keys command:
1. Press “SET” to select
2. Press return key move to next operation mode
3. Up/Down arrow keys to adjust value or select
type
Main Screen: SV = 375
Press return key and release
Operation
Mode
Select
type/value Comment
r-S
Run
Run/Stop
SP
0**
Decimal point position
AL1H
0
Upper-limit alarm 1
LoC
OFF
Lock mode (lock all
keys or only up/down
arrow able to use)
** If using a type-J or type-K thermocouple, SV = 0
If using an RTD, SV = 1
Press and hold down “SET” for 5-sec
Operation
Mode
Select
type/value Comment
InPt
J**
Input type
tPUn
C
Temperature unit
tP-H
800
Upper-limit temperature
range
tP-L
0
Lower-limit temperature
range
Operation
Mode
Select
type/value Comment
CTRL
ON/OFF
Control mode (ON/OFF,
MANUAL, PID and PID
PROG)
S-HC
Heat
Heat/Cool control
ALA1
2
Alarm operates when
PV value is higher than
SV value + ALH setting
SALA
OFF
System alarm
CoSH
ON
Communication write
function that able to use
set point from software
C-Sl
ASCII
Format type
C-no
4
Communication address
bPS
9600
Communication baud
rate
LEn
7
Data length setting
PrtY
Even
Parity bit setting
StoP
1
Stop bit setting
** If using a type-J thermocouple, InPt = J
If using a type-K thermocouple, InPt = K
If using an RTD, InPt = Pt
Press “SET” and release
Operation
Mode
Select
type/value
Comment
HtS
0
Heating hysteresis
setting
tPoF
0
Temperature correction
Revision 02/15/12