beautypg.com

1104 oxygen combustion vessel – Parr Instrument 1104 User Manual

Page 6

background image

1104 Oxygen Combustion Vessel

P a r r I n s t r u m e n t C o m p a n y

6

Preparing the Fuse When the Combustion

Cage is Not Used (Continued)

Set the bomb head on the support stand (P/N A38A)
and attach one end of a 10 cm length of fuse wire
between the two electrodes.

Fasten one end of the wire to the loop electrode
(steps a thru d) then attach the other end to the
straight electrode (steps e thru h). Pull the loop
downward to tighten the connections then bend the
wire upward (step i). Place the capsule in the loop
holder and bend the wire down to the sample (step
j).

Note: Do not submerge the wire in liquid or

powder samples; better combustion will be ob-

tained if the loop of the fuse is set slightly above

the surface. When using a sample that has been

pressed into a pellet, bend the wire so that the

loop is positioned about a centimeter above the

sample.

Closing the Bomb

Close the 1104 Oxygen Combustion Vessel by slid-
ing the head into the cylinder and pushing it gently
until the head is pressed against the inside ledge
of the cylinder. Confirm that the compression ring
(P/N 212A2) is in place above the head. Ensure that
all the compression screws in the screw cap do not
protrude through the underside of the cap. Attach
the screw cap and turn it down by hand as far as it
will go. Progressively tighten the circle of compres-
sion screws in a star pattern until each bears firmly
against the compression ring.

Note: Do not use the spanner wrench (P/N 219A)

to seal the bomb. Extreme force is not required

when tightening the cap. It is important that the

screws be tightened uniformly so that the clos-

ing forces are evenly distributed.

Filling the Bomb Using a Calorimeter

The oxygen inlet connection on the 1104 Oxygen
Combustion Vessel differs from the connection used
on the 1108 Oxygen Combustion Vessel in that it is

Attaching Fuse Wire with Loop Electrode